1.
INTRODUCTION
Sterilization is a process
making the equipments fee from all micro organisms which are not detectable
with standard culturing method There are so many procedures for the
serialization purpose. Some of the important procedures are steam
sterilization, liquid sterilization, dry heat sterilization etc. In the
case of steam sterilization the steam under high pressure is used for the
sterilization purpose.
Under ordinary circumstances
(at standard pressure), liquid water cannot be heated above 100 °C in an open vessel
except for special situations. Further heating results in boiling, but does not
raise the temperature of the liquid water. However, when water is heated in a
sealed vessel such as an autoclave, it is possible to heat liquid water to a
much higher temperature. As the container is heated the pressure rises due to
the constant volume of the container (see the ideal gas law). The boiling point
of the water is raised because the amount of energy needed to form steam
against the higher pressure is increased
The SMART STERIZER is an advanced version of an ordinary
autoclave. The working procedure of this table top sterilizer is similar as
that of autoclave. But here some additional controls are present. That is,
automatic time control temperature control etc.
2.
BASIC PRINCIPLE
The basic principle of steam
sterilization process is that the steam is much heavier than air. So steam
enters the chamber, it fills the upper areas as it is tess dense than air. This
compresses the air to the bottom, forcing it out through a drain. Often a
temperature sensing device is placed in the drain. Only when air evacuation is
complete should discharge stop. Flow is usually controlled through the use of a
steam trap or a solenoid valve, but bleed holes are sometimes used, often in conjuction
with a solenoid valve. As the steam and air mix it is also possible to force
out the mixture from locations in the chamber other than the bottom.
Steam sterilization is
generally carried out at 121°C (250°F) for 15 minutes or at 134°C for 3-4 minutes. Temperatures
can be reduced to 115°C, and even as low as 105°C, depending upon the bi bio
burden integrity, heat resistance, and characteristics of the material being
sterilized. Low-temperature steam processes (65°-80°C) have been used (e.g.,
steam-formaldehyde); however, other combinations could also be used.
Steam (water vapor) is a
ubiquitous compound. Steam delivers high heat condensation, and it is an
activating agent. Before a dormant spore can begin germination and outgrowth,
it must be activated. However, at higher-temperatures, steam becomes
spermicidal. Sterilization, by definition, destroys or eliminates resistant
microbes, including bacterial spores such as anthrax.
More-resistant microorganisms (e.g.,
prions) cannot be eliminated using most standard methods. Extended and
high-steam sterilization, however, can at least reduce the effectiveness of
these organisms. Using the classical definition of sterilization, it is an
absolute criterion. A method has to be capable of destroying or eliminating all
forms of life. In practice, however, Sterilization is best defined as a process that is
capable of delivering a certain probability that an exposed or treated product or material
is free from viable microorganisms, including resistant microbial spores.
3.
BLOCK DIAGRAM
|
PRESSURE
VALVE
|
|
CHAMBER
COIL
|
|
AMPFIER
|
|
SAFTY
TEMPARATURE
LIMIT
|
|
HEATER
ELEMENT
CONTROL
|
|
ADC
|
|
ALARM
VISUAL AUDIO
|
|
SYSETM PROCESSOR
CONTROLLER
|
|
USER
INTERFACE
KEYS
|
|
LM35IC
|
FIG 3.1
4.
BLOCK DIAGRAM DESCRIPTION
The important blocks present
in the case of a table top sterilizer are as explained below:-
CONTROL UNIT
The entire control of this
table top sterilizer relies on a microcontroller 1C (89C51), which is an 8-bit
microcontroller available in 40-pin 1C package. The major sub unites of this
control unit are - Temperature
control:-here we can select appropriate temperature cycle for different objects
to be sterilized. In this case there are two temperature cycles -hot cycle (130
c) and cold cycle (110 c). Time control:- here
we can select proper time periods -sterilization time and dry time. When the
temperature in the chamber reaches to the selected temperature cycle (hot or
cold), the sterilization time will start to count down, when it reaches to zero
dry time will start. When the dry time is over the sterilization process is
completed.
TEMPERATURE
SENSING UNIT
For the accurate measurement
and monitoring of the temperature inside the sterilization chamber, a precise
sensing unit is required .For this purpose a LM35 heat sensor 1C is used which
is placed inside the chamber. LM 35 is a temperature transducer which converts
the temperature into corresponding voltage level.
AMPLIFIER
AND ADC SECTION
The output voltage from the thermocouple corresponding to the temperature
is in the micro voltage range. Hence there is amplifier section Is needed for
further processing of the signal .The section includes a non inverting
amplifier. In order to feed this analog signal to microcontroller it should be
converted into digital form .Hence there is an ADC section is needed.
KEYBOARD
AND DISPLAY SECTION
This section includes four
keys for select the variables and an LCD display is used for the visualization
of these variables.
POWER SUPPLY SECTION
This section will supply
appropriate power to various sections of the circuit Here we use three pin
regulator Ics for this purpose.
CHAMBER AND HEATING COIL
This section consists of completely closed, pressure with stand
able stainless steel chamber with gasket and a 10QOw heating coil.
5. CIRCUIT
DIAGRAM
FIG 5.1
6.
CIRCUIT DESCRIPTION
The
circuit diagrams of the smart sterilizer consist of the following sections:
POWER SUPPLY SECTION
It
consist of a step down transformer (12v), center tapped rectifier, filter
circuits and a regulator 1C The main role of this section is to provide
appropriate power to the various sections. A capacitor filter is used to filter
out the ac components. The regulator 1C used is 7805(5v).
AMPLIFIER
AND ADC SECTION
The
output voltage from the thermocouple corresponding to the temperature is in the
micro voltage range. Hence there is amplifier section is needed for further
processing of the signal The section includes a non -inverting amplifier. An
op-amp CA 3130 is used as the non inverting amplifier. This will provide the
maximum gain.
The signal from the amplifier is analog on nature and it is then
converted into digital form by using an ADC the ADC 0804 is used for this purpose. ADC08O4
The
ADC0804 is CMOS 8-bit successive approximation A/D converters that use a
differential potentiometric ladder similar to the 256R products. These
converters are designed to allow operation with the NSC800 and INSSOSO A
derivative control bus with TRI-STATE output latches directly driving the data
bus. These A/Ds appear like memory locations or I/O ports to the microprocessor
and no interfacing logic is needed. Differential analog voltage inputs allow
increasing the common-mode rejection and offsetting the analog zero input voltage value. In addition, the
voltage reference input can be adjust to allow encoding any smaller analog voltage
span to the full 8 bits of resolution.
7.
CONTROL SECTION
The heart of the control
section is a 89C51 microcontroller 1C which controls and co ordinates the
entire operation of the device. A 6 MHZ crystal is used to drive the
microcontroller .Which is connected in between the 18th and 19th
terminal .An RC network is connected to 9th terminal(RST) of the
microcontroller for the proper working of the microcontroller. For switches and
3 control signals of the LCD is connected to the port 0 of the controller. The
output of ADC is connected to the port 1 of the controller. A 100k resistor
package is connected to the port 0 because of it has no pull up resistor .port
0 is an address/data port. Port 2 is directly connected to the data pins of the
LCD The data from the controller is given to the LCD through this 8 lines of
port 2. port 3 (2,3,4) is used for giving the control signals to the relay and
buzzer driver sections.
RELAY AND BUZZER DRIVER SECTION
A single contact 6 A relay
is used here for provide constant supply to the heater element. Two NPN
transistors (BC 547, TIP 122) are used for drive the relay. The buzzer driver
section includes A 10 k resistor and a PNP transistor (BC 557). This will drive
the buzzer.
LM35 PRECISION CENTIGRADE TEMPERATURE SENSORS
TheLM35 series are precision
integrated-linear temperature sensors calibrated in * Kelvin, as the user is
not required to subtract a large constant voltage from its output to obtain
convenient Centigrade scaling. TheLM35 does not require any external
calibration or trimming to provide typical accuracies of ±WC
at room temperature and ±34°C over a full -55 to
+150°C temperature range. Low cost is assured by trimming and calibration at
the wafer level.The LM35'slow output impedance, linear output, and precise
inherent calibration make interfacing to readout or control circuitry especially
easy.
8.
WORKING
The power supply section
will generate the 12V and 5V then supply to the various sections. Here we use 1
regulator Ic, when the output voltage coming from the secondary winding of the
transformer is rectified and filtered, then given to the first pin of the
regulator Ic. The Ic regulates the voltage into appropriate level.
The voltage output of the LM 35 IC is
amplified by using a non inverting amplifier (CA 3130). This will provide the
maximum gain (A= 1+Rf/R1). An RC network is connected to the non inverting
terminal of OPAMP, which is a low pass filter. For the gain adjustment a 100 k
preset is connected in between the output and inverting terminal. A 10 K preset
is used for adjust the offset voltage. The amplified signal is then feed to the
non inverting terminal of the summing amplifier for make the accurate value.
Then the output voltage of the summing amplifier is inverted in nature hence
the inverting amplifier is used here for adjust the polarity.
The signal from the amplifier
is analog on nature and it is then converted into digital form by using an ADC
.The ADC 0804 is used for this purpose. The successive approximation method is
used here for the conversion. The output of the amplifier section is feed to
the 6th pin of the ADC. The digital output is obtained in the 11th
to 18th pin of ADC, which is then given to the microcontroller for
further processing. The 89C51 is an 8 bit microcontroller which works under the
base of a program. When microcontroller executed the various parameters is
displayed on the LCD. The relay driver and buzzer driver sections are works
under the control of microcontroller.
9.
PCB FABRICATION
LAYOUT DESIGN
From the schematic diagram,
the layout is prepared on a graph paper according to the rules and regulations
of PCB layout In the 2:1 scale. If necessary we can acquire the help of a
computer tapes. In double sided PCB's the track side and the component artwork
are prepared on separate sheet using black colored tapes. In computerized PCB designing,
suitable software is used to design the PCB's. here we use the computerized PCB
designing with the help of pen plotter the artwork is transferred to the paper
2:1 scale.
PHOTOGRAPHIC REDUCTION
The artwork is
photographically reduced to 1:1 scale in process camera. In the orthochromatic
lithe film, the artwork is photographed and the negative film is taken for the
PCB fabrication in photographic method.
BANKING OF LAMINATE
Require size of copper clad
sheet is cut from the copper clad laminate. While cutting of laminate 10mm
clearance is provided.
DARK ROOM PROCESS:
PHOTO RESIST COATING
Photo resist is a liquid
formulation, usually bromide compounds, which is very sensitive to UV light, if
causes polymerization and changes its solubility in solvents. The cleaned
copper clad sheath is dipped in an oven for 10 minutes.
EXPOSING
The photo resist coated board is exposed to an UV light source
through negative film obtained by photographic reduction. The time taken for
exposing
is 2.45 minutes. During
exposing film and copper sheaths are intimate contact. Vacuum pump expose pump
expose machine is used for exposing. For getting good quailing PCB after
exposing, we should follow the cleaning procedure strictly. After exposing you
can develop the PCB.
DEVELOPING
During exposure, the photo
resist coated board has undergone selective polymerization. Thus the circuitry
positions where conductors, the resist unaffected. This unaffected resist is
soluble in developer and polymerized resist is insoluble. After this the board
is dipped in photo resist dye and extra coating. Then the board is washed in
running water.
ETCHING
For removing copper from the
non-polymerized part of the bourdon is fed to the etching machine. The chemical
used for etching is ferric chloride and water having specific gravity 1.2 after
3 or 4 minutes this is rinsed in running water. Over etching will case under
cutting. Then the paths are short circuited, we can remove that one. That means
if any two conductor paths are circuited, we can remove that over. CLEANING
The Contaminates like oil, Oxides are removed by
washing the copper clad sheath with 5% Hcl and scrubbed with pumice powder. If
the water is uniform without any brake, then the surface is full of dirt you
can use sandpaper for cleaning and then washed with water.
DRILLING
The holes for mounting the
components are drill machine having a maximum speed of
20,000 rpm. At the time of drilling, the diameter of the drilled holes proportional to the components, which mounted
at that hole. That the drill bit is selected bit is selected according to the
component.
TIN
PLATING
The tin plating is given to
the PCB for preventing the formation of oxide coating. This also use a special
attraction to the component mounted PCB. Instead
of thin we can also use silver or other
materials, which has the same property of this metals, for the planning of
PCB's.
SOLDERING
The components are bent in the correct dimension, inserted in to
the hole and Soldered.
.
10.
PCB LAYOUT
PCB SIDE
FIG 10.1
PROCEDURE
1. Clean the objects to sterilized and place it inside the
sterilization' chamber.
2. Close the gasket properly.
3. Switch on the main supply.
4. Select the temperature cycle by using H or C keys.
5. Set the sterilization time and dry time.
6. Press the menu button for starting sterilization.
7. Wait until the buzzer is activated.
8. Open the gasket and move out the sterilized objects.
11. HARD WARE AND SOFTWARE REQUIREMENTS
COMPONENT NUMBER OF
ITEMS
RESISTOR 6
RESISTOR ARRAY 1
TRANSFORMER 1
PRESET 2
CAPACITOR 10
2-PIN POWER CORD 1
20PINBASE 1
DIODE 2
ZENER 1
TRANSISTOR 2
CRYSTAL 1
RELAY 1
MICROCONTROLAR 1 C 1
REGULATOR 1 C 1
OP-AMP 1
ADC 1
LCD 1
LM35 1
BUZZER 1
SWITCHES 1
12.
STERILIZATION TIMINGS
·
Laundry: 250°F (121°C) for a minimum
of 30 min.
·
Trash: 250°F (121°C) for at least
45 minutes per bag. Size of the autoclave and of the bags greatly effect sterilization time. Large bags in a small autoclave may require
90 minutes or more.
·
Glassware: 250°F (121°C) for a minimum
of 25 min.
·
Liquids: 250°F (121°C) for 25
minutes for each gallon.
·
Animals & bedding: Steam
autoclaving not recommended (sterilization time required would be at least 8
hours). Incineration in an approved facility
is the recommended treatment of these wastes.
13.
PRECAUTIONS
1.
Ensure that the power plug
is connected properly.
2.
The object to be sterilized
should withstand the temperature inside the chamber.
3.
The gasket should be
properly closed to avoid leakage of steam and losing of pressure.
4.
The gasket should not be
opened while the process is going on as a high pressure is present inside.
5.
Ensure the water level is
above the coil before sterilizing and never switch on the sterilizer if water
is not present which may damage the coil.
6.
The sterilized objects
should be removed with utter care, which should not be touched by hand.
7.
Switch off the device after
the sterilization is complete, do not leave the device powered.
14.
FLOW CHART
15. ADVANTAGES
·
We
can easily operate this instrument.
There is no need of external
Power to drive it. It also turned ON and OFF
automatically.
·
Highly
reliable and small in size.
·
It
is highly sensitive as compared to the
normal sterilizer.
16.
DISADVANTAGES
·
If any fault occurs to the
system, maintenance cost would be high.
·
Trouble shooting is
difficult.
17. CONCLUSION
Our sterilizer is
an automatic device which is controlled by a microcontroller 1C. So the
operator will just have to switch on the device and set the sterilization time
and dry time and the mode in which it is to be operated, i.e. in hot cycle or
cold cycle. When the sterilization procedure gets completed the display
indicates the ending of sterilization and a buzzer is activated, so there is no
need of continues verification of the process.
Our project tabletop
sterilizer has been made to a success by the restless efforts and hard works of
the project team members and with the whole hearted support from all the staff
and students of the institution
18.
FUTURE SCOPE
By using this
smart sterilizer we can sterilize the medical equipments automatically and
these are done in an automatic manner. Its main objective is to avoid the presence
of micro organism we know that, due to micro organism there may arise syndroms.
These an be prevented by using this system.
APPENDIX
MICROCONTROLLER PROGRAM
#include <reg51.h>
#define uc unsigned char
unsigned int tempdata,settemp,temp;
void lcd_com(uc command_word);
void lcd_data(uc value);
void display(uc *s);
void lcd_init();
sbit rs=P1^6;
sbit en=P1^5;
sbit start=P1^1;
sbit oe=P1^0;
sbit eoc=P1^2;
sbit inc=P0^1;
sbit ok=P0^2;
sbit cancel=P0^3;
sbit dec=P0^4;
sbit heater=P1^3;
sbit buzzer=P1^7;
void delay(unsigned int y);
void converts(unsigned int g);
void main()
{
P3=0xff;
P2=0;
P0=0xff;
lcd_init();
lcd_com(0x80);
display("Smart Sterilizer");
lcd_com(0xc0);
display("Set Temp:");
//while(1)
//{
// led=0;
// delay(500);
// led=1;
// delay(500);
//}
while(1)
{
if(ok==0)
{
lcd_com(0xc0);
display("Set
Temp: ");
while(ok==0);
delay(100);
while(ok==1 &&
cancel==1)
{
if(inc==0)
settemp++;
if(dec==0)
settemp--;
if(settemp>=100
|| settemp<=10)
settemp=28;
converts(settemp);
delay(300);
}
}
oe=1;
eoc=1 ;
start=1; //start of convertion
delay(1);
start=0;
delay(5);
start=1;
while(eoc==1);
oe=0; //read data
delay(5);
tempdata=P3;
delay(5);
oe=1;
temp=(tempdata+100)/4;
converts(temp);
//converts(tempdata);
if(tempdata>=settemp)
{
buzzer=0;
heater=1;
delay(400);
}
else
{
buzzer=1;
heater=0;
}
}
}
// lcd..............................
void lcd_init()
{
lcd_com(0x38);
delay(1);
lcd_com(0x0c);
delay(1);
lcd_com(0x01);
delay(1);
lcd_com(0x02);
delay(1);
return;
}
void lcd_com(uc command_word)
{
rs=0;
P2=command_word;
en=0;
delay(1);
en=1;
delay(1);
return;
}
void lcd_data(uc value)
{
rs=1;
P2=value;
en=0;
delay(1);
en=1;
delay(1);
return;
}
void display(uc *s)
{
while (*s)
{
lcd_data(*s);
delay(1);
s++;
}
return;
}
///////convert..............
void converts(unsigned int g)
{
unsigned int x;
x=g/100;
x=x+(0x30);
lcd_com(0xcc);
lcd_data(x);
g=g%100;
x=g/10;
x=x+(0x30);
lcd_com(0xcd);
lcd_data(x);
x=g%10;
x=x+(0x30);
lcd_com(0xce);
lcd_data(x);
return;
}
// delay.............................
void delay(unsigned int y)
{
unsigned int k,h;
for(h=0;h<y;h++)
for(k=0;k<=250;k++);
return;
}
No comments:
Post a Comment